RJ683X AC and DC voltage insulation tester, with AC voltage, DC voltage, insulation resistance test functions, for battery modules, pack, new energy vehicles, household appliances, power electronics and other industry products to provide fast and efficient safety combination test.
n Meet the test requirements of CCC, IEC, EN, VDE, BS, UL, JIS and other mandatory electrical safety standards;
n Ac withstand voltage, DC withstand voltage, insulation resistance items automatic connection test;
n Arc detection, arc detection function, while product test arc alarm output lock, to prevent continuous arc discharge product burn caused by irreparable damage;
n Discharge function, DC voltage resistance or insulation resistance project after the test set discharge project, according to the capacitive size set discharge time, display discharge process voltage, intuitive understanding of discharge results;
n 30 test groups, each group of 8 test steps, easy to operate;
n RS232 interface, PLC interface.
Selection Guide
Features Model number | Ac withstand voltage (ACW) | Dc withstand voltage (DCW) | Insulation (IR) |
RJ6836H | Square root | Square root | Square root |
RJ6835H | Square root | Square root | Square root |
RJ6833H | Square root | Square root | x |
Model number | RJ6836H | RJ6835H | RJ6833H |
Ac withstand pressure test | Output capacity | 500VA (5000V/100mA) | 200VA (5000V/40mA) |
Output voltage range/resolution/accuracy | 50V~5000V /1V / ±(2%rdg.+5dgt.) |
Voltage frequency range/accuracy | 50Hz / 60Hz / ±0.1% |
Current lower limit setting range | 0.00mA~99.99mA | 0.10mA~40.00mA |
Current upper limit setting range | 0.10mA~99.99mA | 0.10mA~40.00mA |
Time setting range/resolution/accuracy | 1.0s~999.9s / 0.1s/±(0.2%setting+2dgt.) |
Current measurement range/resolution/accuracy | 0.10mA~99.99mA 0.01mA ±(2%rdg.+5dgt.) | 0.10mA~40.00mA 0.01mA ±(2%rdg.+5dgt.) |
Dc withstand voltage test | Output capacity | 60VA(6000V/10mA) |
Output voltage range/resolution/accuracy | 50V~6000Vdc / 1V / ±(3%rdg.+10dgt.) |
Current lower limit setting range | 0mA~9.999mA |
Current upper limit setting range | 0mA~9.999mA |
Time setting range/resolution/accuracy | 1.0s~999.9s / 0.1s / ±(0.2%setting+2dgt.) |
Current measurement range/accuracy | 0mA~9.999mA / ±(2%rdg.+0.2%f.s.(20dgt.)) |
Insulation resistance test | Output voltage range/resolution/accuracy | 25V~3000V/1V / ±(2%rdg.+5dgt.) |
|
Resistance lower limit setting range | 0.3MΩ~99.9MΩ/100MΩ-50000MΩ |
|
Resistance upper limit setting range | 0~50000MΩ |
|
Time setting range/resolution/precision | 1.0s~999.9s/0.1s/±(0.2%setting+2dgt.) |
|
Resistance measurement range/accuracy | 25V-199V: 0.3M-99.9M 3%rdg.+3dgt. 100-500M 5%rdg.+5dgt. 200V-299V: 0.3M-99.9M-199M 3%rdg.+3dgt. 200M-4999M 8%rdg.+2dgt. 300V-3000V: 1M-99.9M-199M 2%rdg.+3dgt. 200M-4999M 5%rdg.+5dgt. 5000M-9999M 8%rdg.+2dgt. 10000M-50000M 15%rdg.+30dgt. |
|
Other indicators | External interfaces | RS232,PLC |
Use environment | 0℃~40℃,20%RH~70%RH |
Power supply | AC220V,50/60Hz |
Heavy weight | 17.5 kg |
Exterior dimensions | 342mm×133mm×460mm (W×H×D) without foot, foot height 15mm |