Home>Products>New energy testing instrument>RJ3545 Precision DC Resistance Tester

RJ3545 Precision DC Resistance Tester

Product introduction Main Features Technical Specifications Download

RJ3545 Precision DC Resistance Tester can measure the resistance of motors, transformers and other coils and the contact resistance of relay contacts, and measure the DC resistance of fuses, resistors and various conductive materials through the four-terminal method. Built-in OVC voltage compensation function makes the test results more accurate.

RJ3545 not only meets the requirements of high-precision testing in the laboratory, but also has a variety of communication interfaces to meet the needs of high-volume testing of automatic production lines.


Wide measuring range 0.01μΩ-1200MΩ, 12 steps, max. 6.5 display digits;

Maximum test current 1A;

Flexible I/O interface, compatible with various test systems;

OVC thermopotential compensation function, effectively eliminating thermopotential effects;

Low current mode test prevents the object under test from fusing, heating or magnetizing.

Multi-channel scanning mode can be set, external RJ35-SXX multi-channel scanning unit (optional), to realize multi-channel test.

 


Model

RJ3545

Measuring range

0.01μΩ-1200ΜΩ   12 level

Measuring current

0.1μA1A

Open Voltage

DC 20V (range 10 kΩ1000MΩ)

Maximum 5.5 V (range 10mΩ1000Ω

Measuring

 range

Large current mode

10.00000mΩ-1000MΩ

Large current OVC pulse mode

 

10.00000mΩ-1000Ω

Small current mode

 

100.0000mΩ-100Ω

Large current OVC pulse  mode

 

100.0000mΩ-100Ω

Low current OVC pulse mode

 

1000.000mΩ-1000Ω

Measuring Accuracy

Large current mode

± 0.06%rdg.+0.08%f.s.) 10mΩ

± 0.06%rdg.+0.01%f.s.) 100mΩ

± 0.012%rdg.+0.008%f.s.) 1000mΩ

± 0.008%rdg.+0.008%f.s.) 10Ω

± 0.007%rdg.+0.001%f.s.) 100Ω

± 0.006%rdg.+0.001%f.s.) 1000Ω

± 0.007%rdg.+0.001%f.s.) 10kΩ

± 0.007%rdg.+0.001%f.s.) 100kΩ

± 0.008%rdg.+0.2%f.s.) 1000kΩ

± 0.03%rdg.+0.2%f.s.) 10MΩ

± 0.2%rdg.+1%f.s.) 100MΩ100M High-precision mode

Below 10.00M :±0.5%rdg.+0.02%f.s.) 100MΩ100M Low-precision mode

Above 10.01M :±1%rdg.+0.02%f.s.) 100MΩ100M Low-precision mode

Below 100.0M:±1%rdg.+0.02%f.s.) 1000MΩ100M Low-precision mode

Above 100.1M :±10%rdg.+0.02%f.s.) 1000MΩ100M Low-precision mode

Large current OVC pulse mode

± 0.06%rdg.+0.001%f.s.) 10mΩ

± 0.06%rdg.+0.001%f.s.) 100mΩ

± 0.012%rdg.+0.001%f.s.) 1000mΩ

± 0.008%rdg.+0.005%f.s.) 10Ω

± 0.007%rdg.+0.001%f.s.) 100Ω

± 0.006%rdg.+0.001%f.s.) 1000Ω

Small current mode

± 0.014%rdg.+0.05%f.s.) 100mΩ

± 0.008%rdg.+0.05%f.s.) 1000mΩ

± 0.008%rdg.+0.05%f.s.) 10Ω

± 0.008%rdg.+0.01%f.s.) 100Ω

Large current OVC pulse  mode

± 0.014%rdg.+0.006%f.s.) 100mΩ

± 0.008%rdg.+0.002%f.s.) 1000mΩ

± 0.008%rdg.+0.005%f.s.) 10Ω

± 0.008%rdg.+0.010%f.s.) 100Ω

Low current OVC pulse mode

± 0.2%rdg.+0.003%f.s.) 1000mΩ

± 0.2%rdg.+0.002%f.s.) 10Ω

± 0.2%rdg.+0.002%f.s.) 100Ω

± 0.2%rdg.+0.002%f.s.) 1000Ω

External interface

RS232/485 communications interface, Handler interfacePLC interface

Other Functions

OVC voltage compensation, comparator function, low current test mode, low current test mode, contact detection function, resistivity test function, temperature test (optional), temperature compensation function (optional)

Usage environment

20℃26℃20%RH70%RH

Power supply

AC220V±10%50/60Hz20VA

Weight

3.7Kg

Dimensions

213mm×88mm×350mm(W×H×D) excluding the feet height, the feet height is15mm

Accompanying documents and Accessories

1 instruction manual, 1 power cable (1.5 m), 1 communication cable (1.5 m), 1 set of test cables (1.5 m),1GPIB connector plug


Customer site

Products

Ruijie Instrument

Factory Address: No.38 Longfei Road, Chengyang District, Qingdao, China

4008-515-616

< a href="https://www.ruijie-ate.com/">在线客服